DCC
Models: UCClite, UCClite2, UCC1, UCC2, Fusion, T3, T3+, T5, S3, S3+
Supported Probes: PH20, PH20+, Revo-2, PH10 and all other Renishaw indexable probes with digital probing.
Supporting Probing: TP2, TP20, TP200, SP25, RSP2 & 3 (Revo)
UCC Suite
Due to a change in how UCC Server deals with "blocks" of commands, CMM-Manager (and some other brands of inspection software) will not run properly with SP25, PH20, or REVO.
*** CMM-Manager with UCC Server 5.6.x, 5.7.0, 5.7.1, 5.9.x , or 5.10.0 version is not supported
***
*** UCC Server 5.7.2 - 5.8.6, 5.10.1 and 5.11.3 are acceptable ***
We will not be able to offer assistance with systems running UCC Server 5.6.x or 5.7.0 / 5.7.1 / 5.9.x / 5.10.0 with SP25, PH20, or REVO.
Starting in version 3.7 (UCC Suite 5.0.3+ must be used), Full Sync (X-Path) is supported in CMM-Manager. Users will no longer configure the machine, probes, etc. in CMM-Manager; all machine and probe data will sync to CMM-Manager upon connection to UCC Server via I++ interface. See the attached UCC I++ Server User Guide for details: User Guide - UCC I++ Server.pdf.
Suggested software versions (CMM-M / UCC Suite)
3.6 / 5.0.5 (All)
3.7 / 5.0.5 (PH20) - 5.1.0 (All others)
3.8 / 5.7.2
3.9 / 5.7.2
3.9 SP2 / 5.7.2
3.10 / 5.7.2
2020 / 5.7.2
2021 / 5.7.2
PH20 Head ReQualify can now be done in UCC Server - use command RWReQualify(Head()) from the command tab in UCC Server.
PH20 will exhibit slower performance when running programs in version 3.6. This was fixed in version 3.6 SP1. If you are running 3.6 and have not yet updated to SP1, please download the
IQIppDmeHotFixfor3.6.zip and replace the IQIppDme.dll, which is located "C:\Program Files (x86)\Nikon\CMM-Manager 3.6\Driver\."
3.9 SP2 adds support for "direct connect" for systems that do not utilize SP25/Scanning or PH20. For more information, see
UCC Direct 2.0 (DCC)