Hexagon (B&S) - Identify Sharpe32 vs. Leitz controller

Hexagon (B&S) - Identify Sharpe32 vs. Leitz controller

The Sharpe 32, Shrp32z, Leitz protocol controllers can be packaged in a similar unit and not easily distinguished. Here are several ways to identify Sharpe 32 vs. Leitz protocol:

1. Check communication settings in current software
  1.  Sharpe: COM 1 4800 E 7 1 // com interface (port,baud,parity,data,stop)
  2.  Leitz: COM 1 9600 N 8 1 // com interface (port,baud,parity,data,stop)
2. Check for Downl.oad or Syspar configuration file on existing system
  1.  Sharpe: Either Downl.oad or Syspar (older systems) will exist on the PC (see install guide for location)
  2.  Leitz: All config is stored on controller there should be no config file stored on local PC
3. Use Hyperterminal
  1.  Start Hyperterminal, and press Ctrl+ECB.
  2.  If it is Leitz protocol, it should display “Ready”, and further command can be issued, e.g. “TESTSOFT”
4. Check cable connection
  1. Sharpe Protocol
    1. dongle on "Servo Key" connector
    2. RS232 cable on H1J0 (DB9 female)
    3. RS232 cable on H2J0 for Renishaw PHC 9/10
  2. Leitz Protocol
    1. no dongle
    2. RS232 cable connected to DB9 male connector next to "Servo Key" connector
If system is Sharpe32 see this install guide - Hexagon (B&S) - Sharpe32 (DCC)

If system is Leitz see this install guide - Hexagon (B&S) - Leitz / Sharpe32z / RC1 (DCC)